What Transport Canada published
- Recall number
- 2000056
- Recall date
- 8 March 2000
- Make
- Demers
- Models
- Ambulance
- Model years
- 1992 to 1999
- Vehicle class
- Light Truck & Van
- System
- Other
- Raised as
- Safety TC
- Vehicles affected
- 411
- Register
- Canada — Transport Canada, Vehicle Recalls Database
On certain vehicles, the stainless steel hose that links the oxygen cylinder to the manometer may rupture and explode when the cylinder is opened. This could result in a fire. Correction: Oxygen cylinder will be disconnected from the stainless steel hose and the manometer will be installed directly on the cylinder. Stainless steel hose will be cut to prevent further use.